Learn More About the Accuracy of Our BioProTT™ FlowMeasurement System

Flow Measurement

1. Check the correct sensor positioning on the tube.
Sensor Inlet Section_BIO

2. Find the right sensor position regarding pumps and valves.
TechNote-Accuracy-1

3. Align the sensor and flow path.
TechNote-Accuracy-2

4. Correctly fit the sensor and the tube.
Sensor_Correct Tube Size

5. Check the application parameters and calibration factor in regard to:

  • Medium type and temperature
  • Tube size and material
  • Flow range

6. Before starting the measurement, make sure to:

  • Let the system equilibrate
  • Check the acoustic coupling
  • Carry out a zero flow adjustment